Analysis range |
0-1mmol/L (customizable) |
|
|
■ |
detection limit |
0.01mmol/L |
|
|
■ |
resolution ratio |
0.001mmol/L |
|
|
■ |
error |
±3% |
|
|
■ |
repetitiveness |
≤5% |
|
|
■ |
Range drift |
≤5% |
|
|
■ |
analysis mode |
integral point |
Run by the hour, up to 24 groups per day |
|
margin |
Run 0-9999 minutes at intervals as set |
||
|
hand movement |
Manual analysis |
||
|
long-range |
Remote control by MODBUS; dry contact control |
||
|
■ |
Calibration method |
The interval count |
The interval was calibrated according to the cumulative number of analyses |
|
Number of days between |
The interval is calibrated according to the number of days in operation |
||
|
hand movement |
Manual calibration |
||
|
long-range |
Remote control by MODBUS |
||
|
■ |
Cleaning method |
voluntarily |
Analysis cleaning; fault cleaning; abnormal power failure cleaning; emergency shutdown cleaning |
|
hand movement |
Clean by hand |
||
|
■ |
report to the police |
Low level alarm; Self-test alarm; Fault alarm; Excessive alarm (one relay) |
|
|
■ |
data |
The signal is 4-20mA analog |
|
|
■ |
communication |
An RS232 or RS485 interface standard MODBUS protocol |
|
|
■ |
memory |
About 20,000 groups (automatic power-off save); support USB drive export EXCEL; |
|
|
■ |
put a seal on |
Miniature industrial thermal printer (optional) |
|
|
■ |
Reagent quantity |
About 180 times per set |
|
|
■ |
Maintenance cycle |
About 30 days, 45 minutes |
|
|
■ |
cell display |
7 inch LCD, 800*480 resolution, TFT true color touch screen |
|
|
■ |
service environment |
5-35℃, humidity <90% (no condensation) |
|
|
■ |
Water sample temperature |
5—35℃ |
|
|
■ |
Water harvesting unit |
For water sample circulation and retention; automatic backwash precision filtration system (optional) |
|
|
■ |
size |
73cm×47cm×28cm |
|
|
■ |
weight |
25KG |
|
|
■ |
source |
AC 220V ± 10%, 50Hz ± 1% |
|
|
■ |
power |
150W |
|